What can EARS offer you?
EARS provides you with the ability to quickly confirm that the sound output of your audiometer, as delivered to the earphones directly or through booth connections, is as expected. It also uniquely has a response button testing facility.

How does it do these things?
The earphones are fitted onto the colour coded mouldings, as you would to a person under test, and a cable from Ears is plugged into your audiometer response socket in place of your response button. Then run a full automatic or manual test and compare the thresholds achieved with the target values or previous results recorded.
Your response button can be plugged into the socket marked Response Unit Test on the EARS unit and correct operation confirmed by pressing and releasing the button.

Why check your audiometer?
Checking performance prior to use is recommended by the H&SE in their publication Controlling Noise at Work, in order to pick up degradation in booth sockets/cables, audiometer earphones/cables and audiometer performance.
EARS is ideal for performing this check as it provides accurate and consistent thresholds. USB power provides extra reliability as the device will never stop working during a test due to drained batteries.

The EARS system offers numerous benefits to a user:

  • Simple and Easy To Use
  • Light & Very Portable
  • Fully Automatic System
  • Accurate & Consistent
  • USB Powered

Accuracy: Plus/Minus 10% of reading or 10 L/min flow (whichever is the greater)
Repeatability: Plus/Minus 5%
Measuring: Average of 3 readings
Working Life: Replace after 2 years

  • GM SMART Resuscitation Trainer Light

    SMART Resuscitation Mask Leak Trainer

  • GM Respiratory Flowheads All Sizes

    Respiratory Flowheads

  • Audio Booth

    Audiometer Booths

  • GM EARS Response System

    Electroacoustic Auto Response System